Local Operation
M370430-01 Rev F 3-11
3
Output enabled
When the output is turned on, both the voltage and current rise as quickly
as possible to try and reach the set points (V
SET
and I
SET
) that were
configured prior to enabling the output. Initially the unit will operate in
CV mode, which is the default mode. As the voltage and current rise to
meet the requirements of the load, the set point that is reached first
(based on the R
I
, either V
SET
or I
SET
) will determine which mode the
power supply will remain in. The operating mode is indicated by either
the CC mode LED or the CV mode LED, one of which will illuminate on
the front panel.