TR-2100 User Manual 63
Appendix I - TR-2100 Specifications
Inputs
No. of channels 8, 16, 24 or 32 Analogue
16 - 160 Digital
Unlimited number of channels by cross-triggering
systems
Voltage inputs 57 – 120V RMS typical
Other options available
Current inputs 1A or 5A RMS typical
Inputs via shunts or CTs
CT phase lag correction
Over range 212V RMS (voltage)
Selectable for current (x1 - x20)
Input burden 0.06 VA (voltage)
<0.014 VA (current)
Transducer inputs Any analogue input may be scaled for standard
transducers to record AC or DC quantities
Frequency response DC - 1/2 sampling rate, +0dB, -3dB.
Accuracy Better than 0.1% of full scale
Inputs are temperature compensated and can be auto-
calibrated (no pots required)
Digital inputs 24 / 48 / 125 / 250 Vdc
Contacts settable as normally open or closed
Digital debounce 1 - 10ms in 1ms increments