METHOD OF MEASUREMENT:
ELECTROCHEMICAL OXYGEN SENSORS
The MODEL 210BX utilizes P Series Oxygen Sensors that feature Proprietary Sensor
Technology. We manufacture these electrochemical sensors in our state-of-the-art factory in Costa
Mesa, CA using an innovative design, which enables them to achieve the fastest 'come-down' time in the
industry and swift response times (both upscale and downscale).
Notes:
Any sensor can be used in O
2
applications above its recommended operating ranges, however, it may
shorten the sensor's lifespan.
The minimum range of the sensor is dependent on the sensor as well as the AMI Analyzer it is used within.
Sensitivity is the minimum change in O
2
in the gas stream that will be detected by the sensor.
Notify the factory for use in CO
2
background gas above 50%. The AMI O
2
Analyzer will require
additional programming.
AMI Model Number P-2
Type PERCENT
Recommended O
2
Measurement Range
0 —50%
Minimum Range 0 —1,000 ppm
Sensitivity 10 ppm
Special Conditions
O
2
in
inert gas
0 —25%
Up to
100% CO
2
PERCENT
P-3
0 —1,000 ppm
100 ppm
0 —100%
Enriched
Oxygen
PERCENT
P-4
0 —1.0%
1,000 ppm
0 —25%
<500 ppm H
2
S
100% CO
2
PERCENT
P-5
0 —1,000 ppm
100 ppm
3