No Output Alarm Indication
Resolution(s):
• Verify that the alarm and alarm delay setpoints are correct
• Confirm the Alarm Delay or Alarm Hold Off setting is correct
• Check that the output wires are properly stripped and connected at their correct positions at
their respective terminals
• Verify that the alarms on the Analyzer are properly configured using the COMMAND
CENTER Software (see pages 34 – 37)
Incorrect Readings
Resolution(s):
• Verify that there are no leaks at any gas connections using SNOOP
®
(see page 19)
• Perform an air calibration (see page 24)
Display Pressure Reading Not Correct
Resolution(s):
• Perform the Initiation of the Pressure Sensor Procedure on page 15
Analyzer Refuses to Accept Front Panel Settings
Resolution(s):
• Use the COMMAND CENTER Software to verify that the Security Settings match your
preference
Analyzer Reads Zero
Resolution(s):
• Check that the oxygen sensor is in the correct position and not upside down. If it is, re-orientate
in the correct position
No Voltage or Current Output to Recording Device
Resolution(s):
• Check that the output wires are properly stripped and connected at their correct positions at
their respective terminals
'Err' Flashes on the LCD
Resolution(s):
• Look up the Error Code on page 45 and troubleshoot/resolve it
• If you cannot resolve, contact AMI for further
47