No Output Alarm Indication
Resolution(s):
Verify that the alarm and alarm delay setpoints are correct
Confirm the Alarm Delay or Alarm Hold Off setting is correct
Check that the output wires are properly stripped and connected at their correct positions at
their respective terminals
Verify that the alarms on the Analyzer are properly configured using the COMMAND
CENTER Software (see pages 34 – 37)
Incorrect Readings
Resolution(s):
Verify that there are no leaks at any gas connections using SNOOP
®
(see page 19)
Perform an air calibration (see page 24)
Display Pressure Reading Not Correct
Resolution(s):
Perform the Initiation of the Pressure Sensor Procedure on page 15
Analyzer Refuses to Accept Front Panel Settings
Resolution(s):
Use the COMMAND CENTER Software to verify that the Security Settings match your
preference
Analyzer Reads Zero
Resolution(s):
Check that the oxygen sensor is in the correct position and not upside down. If it is, re-orientate
in the correct position
No Voltage or Current Output to Recording Device
Resolution(s):
Check that the output wires are properly stripped and connected at their correct positions at
their respective terminals
'Err' Flashes on the LCD
Resolution(s):
Look up the Error Code on page 45 and troubleshoot/resolve it
If you cannot resolve, contact AMI for further