3www.amico.com
Vacuum Ports: 1/8" NPT Female
Modes: REG (Regulated) – provides an adjustable, continuous
vacuum level
OFF – no vacuum
INT (Intermittent) – provides an adjustable vacuum level
that cycles between
REG and OFF
FULL - provides full suction
Intermittent Cycle Time: Factory set at approximately 16 seconds ON and 8
seconds OFF (for reference only). All units come pre-set
in the OFF setting.
Operating Environmental Limits: 0°F to 122°F (-18°C to 50°C)
Recommended Environmental
Operating Limits: 55°F to 85°F (13°C to 29°C)
Storage Environmental Limits: -4°F to 140°F (-20°C to 60°C). Max 95%, Non-condensing
Standard Flow Rate: 0 - 70 LPM
Pediatric Flow Rate: 0 - 40 LPM
Remove the Amico Patient Care Corporation Vacuum Regulator from the packaging and inspect it
for damage. If there is any damage, DO NOT USE and contact your provider.
INTENDED USE
WARNING: DO NOT use this Vacuum Regulator for anything other than its intended use.
Personal injury and/or damage to the Vacuum Regulator may result from misuse.
WARNING: Only personnel instructed and trained in its use should operate this Vacuum
Regulator.
SPECIFICATIONS
Vacuum Regulator Model Gauge
Gauge Accuracy
Analog Digital
Continuous / Intermittent 0 – 300 mmHg ± 3% F.S.* ± 1% F.S.*
Continuous High 0 – 750 mmHg ± 3% F.S.* ± 1% F.S.*
Pediatric Continuous Intermittent 0 – 160 mmHg ± 3% F.S.* ± 1% F.S.*
Neonatal Continuous Intermittent 0 – 100 mmHg ± 3% F.S.* ± 1% F.S.*
*F.S. = Full Scale
Receiving Inspection