EasyManua.ls Logo

AmpliVox 270+ - EMC Emissions and Immunity; Electromagnetic Emissions Guidance

AmpliVox 270+
63 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
47
D-0126050 rev 1 - 270+ Operating Manual
7. EMC GUIDANCE & MANUFACTURER’S DECLARATION
This instrument is suitable in hospital environments except for near active HF surgical equipment and
RF shielded rooms of systems for magnetic resonance imaging, where the intensity of electromagnetic
disturbance is high
Use of this instrument adjacent to or stacked with other equipment should be avoided because it could
result in improper operation. If such use is necessary, this instrument and the other equipment should
be observed to verify that they are operating normally
Use of accessories, transducers and cables other than those specified or provided by the manufacturer
of this equipment could result in increased electromagnetic emissions or decreased electromagnetic
immunity of this equipment and result in improper operation. The list of accessories, transducers and
cables can be found in this appendix.
Portable RF communications equipment (including peripherals such as antenna cables and external
antennas) should be used no closer than 30 cm (12 inches) to any part of this instrument, including
cables specified by the manufacturer. Otherwise, degradation of the performance of this equipment
could result
NOTICE
ESSENTIAL PERFORMANCE for this instrument is defined by the manufacturer as:
This instrument does not have an ESSENTIAL PERFORMANCE absence or loss of ESSENTIAL
PERFORMANCE cannot lead to any unacceptable immediate risk
Final diagnosis shall always be based on clinical knowledge There are no deviations from the collateral
standard and allowances uses
This instrument is in compliance with IEC60601-1-2:2014, emission class B group 1
NOTICE: There are no deviations from the collateral standard and allowances uses
NOTICE: All necessary instruction for maintaining compliance with regard to EMC can be found in the
general maintenance section in this instruction. No further steps required.
CAUTION

Table of Contents

Related product manuals