EasyManua.ls Logo

Amprobe AM-560 - Temperature Measurement; Maintenance

Amprobe AM-560
33 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
22
11. Temperature Measurement
Range Resolution Accuracy
-40 – 40°C
1°C
±(2%+8 LSD)
>40 – 400°C ±(1%+8 LSD)
>400 – 1000°C ±2.5%
-40 – 104°F
2°F
±(2%+12 LSD)
>104 – 752°F ±(1%+12 LSD)
>752 – 1832°F ±2.5%
Overload protection: 1000Vp
K-type (nickel-chromium/nichrosi) thermocouple must be used for temperature
measurements.
MAINTENANCE AND REPAIR
If the Meter fails to operate, check battery, test leads, etc., and replace as
necessary.
Double check the following:
1. Replace the fuse or battery if the meter does not work.
2. Review the operating instructions for possible mistakes in operating
procedure.
Quick check on 0.5A FUSE:
Step 1: Turn the rotary switch to mA
position.
Step 2: Use a multimeter with continuity function to verify the fuse
continuity for the fuse of mA/μA terminal. Connect the test leads to mA/μA
terminal and COM terminal.
Continuity buzzer activates: the fuse is OK
Continuity buzzer is not activated: the fuse is burnt. Replace the fuse as specified.
F1 0.5A H 1000V fast-fuse, ( 6.3×32)mm
Quick check on 10A FUSE: