USER MANUAL DENTAL UNIT - QS4-451-01 / SD 300 Scandinavian
20
DISPLAY VIEW
ERROR TYPE
1 No micromotor(electric), or circuit broken, pole-1
2 No micromotor(electric), or circuit broken, pole-2
3 No micromotor(electric), or circuit broken, pole-3
4 Micromotor(electric) short-circuit or overload, pole-1
5 Micromotor(electric) short-circuit or overload, pole-2
6 Micromotor(electric) short-circuit or overload, pole-3
7 Light bulb missing or blown, instrument-pole-1
8 Light bulb missing or blown, instrument-pole-2
9 Light bulb missing or blown, instrument-pole-3
10 Light bulb missing or blown, instrument-pole-4
11 Bulb short-circuit or overload, instrument-pole-1
12 Bulb short-circuit or overload, instrument-pole-2
13 Bulb short-circuit or overload, instrument-pole-3
14 Bulb short-circuit or overload, instrument-pole-4
15 Cannula support safety feature unactive. SYSTEM BLOCKED
16 Failed start-up test 1, source micromotor supply.
17 Failed start-up test 2, source micromotor supply.
18 Failed start-up test 1, source light power supply.
19 Failed start-up test 2, source light power supply..
20 Communications bus failure. SYSTEM BLOCKED
21 Pedal test failure, no response. SYSTEM BLOCKED.
22 Water unit test failure, no response .SYSTEM BLOCKED.
23 Circuit box failure, no response. SYSTEM BLOCKED
24 Electric centred pedal test failure. SYSTEM BLOCKED.
25 Water unit circuit board failure 9015043.
26 24Vca power supply failure or excess consumption.
27 Undervoltage in 24Vdc line.
28 Overvoltage in 24Vdc line
29 Pedal test failure, out of range, potentiometer incorrect.
30 Without communication between keyboard and instrument tray charts.
31 MX driver faulty, DMX module informs there is an error
32 Failure in communication with MX driver, we cannot communicate with DMX.
33 MX Driver configured as Standard, but the Basic model detected.