USER MANUAL DENTAL UNIT - QS4-451-01 / SD 300 Scandinavian
21
9.- SAFETY MOVEMENTS
- The dental chair includes mechanical safety stops for the chair and backrest, ensuring
correct maneuvering and positioning.
- Motor overload control through integrated heat sensors. Should the heat sensors go off,
wait 15 minutes for them to cool.
- The backrest safety feature, in the event of pressure or knocks against other objects Turing
manual or automatic lowering recline, suspends all chair movement (seat and backrest),
before lifting raising it a few centimeters.
- The base safety feature, in the event of pressure or knocks against other objects Turing
manual or automatic lowering recline, suspends all chair movement (seat and backrest),
before lifting raising it a few centimeters.
- The cannula safety feature, in the event of any pressure against the arm, or knocking against
other objects during manual or automatic lowering recline, suspends all chair movement
(seat and backrest).
- The control pedal safety blocks all chair movements of the chair (seat and backrest)
once the control lever is activated, allowing the user to work in the patient’s oral cavity in
complete safety. If the seat is moving, it can be stopped simply by pressing the lever to
active the safety function. If only instrument has been selected by pulling its hose, pedal is
still operative with respect movements
- Safety on the adjustment pedal does not allow automatic movements, so preventing
unexpected activations due to the sensitivity of the three-dimensional movements button,
except for the memorized return-to-zero position.
- If the chair moves to a memorized location, pressing any directional key on the keypad
(main and auxiliary) safety will stop the unit.
- Monitoring system for error control, inactive safety features or cannula support safety
failure, shown on the touch screen.
When a safety is activated, chair performs a movement in the opposite direction to the one seat
or backrest engines are performing, in order to free the trapped component.
10.- CLEANING AND STERILISING
When cleaning your dental unit, it is essential that you use neutral products. Cleaning products
with high chemical content can damage plastic parts or upholstery. When cleaning, take care
not to wet the equipment too much, as its interior is made up of electrical components. Various
specialists in the dental hygiene industry offer a wide range of cleaning products for achieving
optimum results.
ANCAR thoroughly and systematically cleans your equipment, prior to its dispatch,
with Magnolia Spray ECO-JET1 (Cattani Group).
For cleaning and disinfecting the different parts of the system, please check:
- Section “Cannula Support”. (page 22)
- Appendix 1, “Cattani system and suction”.
- Appendix 2, “Metasys system and suction”.
- Appendix 3, “Cleaning other parts of the unit”.
- Appendix 4, “Durr Dental VS300 suction system“.
- Appendix 5, “3F / 6F Syringe maintenance”.
STERILISING AND AUTOCLAVE.
135º C
Hand instruments should be sterilised in the autoclave at an average temperature
of 135°C, though we recommend following manufacturer instructions for best
results.
Cannulas and cannula-holder can also be sterilised. Check the section entitled
“Cannula support”.
Proper use and maintenance of the equipment will prolong its working life.