USER MANUAL DENTAL UNIT - QS4-451-01 / SD 300 Scandinavian
23
APPENDIX 1. CLEANING / DISINFECTING OF CATTANI
GLASS, CANNULAS AND BASIN
Before removing the water unit cover or extracting the vessel, disconnect the
dental unit from the mains by turning off the mains. Use suitable protective
gloves.
As shown in Figure 14, remove the outer cover of the water unit (A), by first pulling lightly
towards you and then lifting it.
Note that this should be gripped from the top using 2 magnets (B) and below using the supports
(C). You will find the vessel installed (D).
Vessel, once at the end of each day
To remove (G), switch off (I), pull down the two rings (F). Grip
the cover (E) lift it slightly and pull towards you from the vessel
(G). Also clean the support (H). Replace the vessel following the
instructions in reverse. Use the Puli-Jet (Cattani) and water.
Cannula circuit, at lunch time and at night.
Cleaning and disinfecting: use Cattani Puli-Jet.
Concentration: 50 cl per liter of water. Take in water through each
cannula.
External cannula-surfaces, after each treatment
Disinfectant: Eco-Jet Cattani.
Rinse basin with water after each treatment.
Cannula filter, foamless disinfectant
Use anti-odour tablets for the cannula filters
Fig. 15 (I) Quick connector.
Caution
Do not clean with foam. Do not use sodium hypochlorite.
'
%
&
$
Fig. 14
E
I
G
F
F
H