EasyManua.ls Logo

Anritsu 681XXC - Initiating Master-Slave Operation; Master-Slave Operation

Anritsu 681XXC
241 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Step 4 Connect the Master instrument RF OUT
-
PUT and the Slave instrument RF OUT
-
PUT to the appropriate connections on the
DUT.
Initiating
Master-Slave
Operation
The following paragraphs describe how to set up
both instruments to perform master-slave opera
-
tions. Use the CW Frequency Mode menu map
(Chapter 4, Figure 4-2) to follow the menu se
-
quences.
To initiate master-slave operation, turn on both
instruments and place them in CW mode. The CW
Menu (below) is displayed.
On the Master instrument, press Master Slave to
go to the Master-Slave Menu display (below).
This menu lets you perform the following:
q
Go to the Slave Frequencies List menu.
q
Set the dF frequency for the Slave unit.
q
Set the Slave unit’s main power level (L1).
q
Set the alternate sweep power level (L2) for
the Slave unit.
q
Turn master-slave operation on and off.
Press Slave Freqs to go to the Slave Frequencies
List Menu display (next page).
681XXC OM 7-5
USE WITH MASTER-SLAVE
OTHER INSTRUMENTS OPERATION
NOTE
Master-slaveoperationsarealways
initiated in the CW frequency
mode. Once initiated, you then can
change to a sweep frequency mode
of operation by selecting the de-
siredfrequencymodeontheMaster
instrument.

Table of Contents