6-3 :CONFigure Subsystem Chapter 6 — CDMA Commands
6-4 PN: 10580-00341 Rev. C MS2720T PM
:CONFigure:OTA <char>
Title: Configure Over-The-Air Measurement
Description: Valid parameter options: PSCAn|MPATh. This command configures
the selected Over The Air measurement. It disables any other active
measurements. Sets the instrument to single sweep mode
(:INITiate:CONTinuous OFF). Measurement settings can be modified
before initiating a measurement. When the PSCAn option is selected,
the measurement results are equivalent to the results that are
displayed by choosing Measurements, then OTA, then Pilot Scan from
the front panel. When the MPATh option is selected, the measurement
results are equivalent to the results that are displayed by choosing
Measurements, then OTA, then Multipath from the front panel.
Parameter: <char>
:CONFigure:RF <char>
Title: Configure RF Measurement
Description: Valid parameter options: SUMMary|SPECtrum|ACPR|EMISsion
This command configures the selected RF measurement. It disables any
other active measurements. Sets the instrument to single sweep mode
(:INITiate:CONTinuous OFF). Measurement settings can be modified
by using the [:SENSe]:RF commands before initiating a measurement.
When the SUMMary option is selected, the measurement results are
equivalent to the results that are displayed by choosing Measurements,
then RF, then RF Summary from the front panel. When the SPECtrum
option is selected, the measurement results are equivalent to the results
that are displayed by choosing Measurements, then RF, then Channel
Spectrum from the front panel. When the ACPR option is selected, the
measurement results are equivalent to the results that are displayed by
choosing Measurements, then RF, then ACPR from the front panel.
When the EMISsion option is selected, the measurement results are
equivalent to the results that are displayed by choosing Measurements,
then RF Measurements, then Spurious Emission from the front panel.
Parameter: <char>
Related Command: [:SENSe]:FREQuency
ООО "Техэнком" Контрольно-измерительные приборы и оборудование www.tehencom.com