9-3 :DISPlay Subsystem Chapter 9 — TDSCDMA Commands
9-4 PN: 10580-00341 Rev. C MS2720T PM
9-3 :DISPlay Subsystem
This subsystem provides commands that modify the display of data for the user. They do not
modify the way in which data are returned to the controller.
:DISPlay:WINDow:TRACe:Y[:SCALe]:OFFSet <rel ampl>
:DISPlay:WINDow:TRACe:Y[:SCALe]:OFFSet?
Title: Power Offset
Description: Sets the power offset value for the y-axis.
Parameter: <rel ampl>
Default Value: 0dB
Default Unit: dB
Range: –100 dB to +100 dB
Front Panel
Access: Amplitude, Power Offset
:DISPlay:WINDow:TRACe:Y[:SCALe]:PDIVision <rel ampl>
:DISPlay:WINDow:TRACe:Y[:SCALe]:PDIVision?
Title: Scale
Description: Sets the scale (dB/division) for the y-axis.
Parameter: <rel ampl>
Default Value: 10 dB/div
Default Unit: dB
Range: 1 dB to 15 dB
Front Panel
Access: Amplitude, Scale/div