6-8 :MEASure Subsystem Chapter 6 — CDMA Commands
6-12 PN: 10580-00341 Rev. C MS2720T PM
:MEASure:OTA:MPATh?
Title: Measure OTA Multipath
Description: Sets the active measurement to Multipath, sets the default
measurement parameters, triggers a new measurement and returns six
Ec/Io values, six Tau values, Channel Power, and Multipath Power. It is
a combination of the following commands :CONFigure:OTA MPATh and
:READ:OTA:MPATh? To make an OTA Multipath measurement with
settings other than the default values send:
:CONFigure:OTA MPATh
Commands to set desired settings
:READ:OTA:MPATh?
Data is returned as 14 comma-delimited values: six Ec/Io Values in dB,
six Tau values in ps, Channel Power in the current user set unit
(dBm/W/mW), and Multipath Power in dB. All invalid values are
replaced with “--”.
Related Command: :CONFigure:OTA MPATh
:READ:OTA:MPATh?
:FETCh:OTA:MPATh?
Front Panel
Access: Measurements, Over-the-Air, Multipath
:MEASure:PFail?
Title: Measure PASS/FAIL
Description: Sets the active measurement to PASS/FAIL, sets the default
measurement parameters, triggers a new measurement and returns the
PASS/FAIL results. It is a combination of the commands :CONFigure
PFail and :READ:PFail? To make a channel power measurement with
settings other than the default values send:
:CONFigure PFail
Commands to set desired settings
:READ:PFail?
Related Command: :CONFigure PFail
READ:PFail?
FETCh:PFail?
Front Panel
Access: Measurements, Pass/Fail Mode