Chapter 6 — CDMA Commands 6-3 :CONFigure Subsystem
MS2720T PM PN: 10580-00341 Rev. C 6-3
6-3 :CONFigure Subsystem
This set of commands prepares the instrument for the selected measurement. It disables any
currently-enabled measurements and activates the specified measurement. It sets the
instrument to single sweep mode, waiting for an :INITiate command. It will not initiate the
taking of a measurement.
Current instrument settings may be changed to default values. These changes are identified
with their respective measurement commands.
:CONFigure SUMMary|PFail
Title: Configure Summary and Pass/Fail Measurements
Description: This command configures the summary and PASS/FAIL measurements.
It disables any other active measurements. None of the instrument
parameters are changed by the execution of this command. Sets the
instrument to single sweep mode (:INITiate:CONTinuous OFF).
Measurement settings can be modified by using the [:SENSe]
commands before initiating a measurement.
Parameter: SUMMary|PFail
Parameter Type: <char>
Front Panel
Access: Measurements, CDMA Summary or Pass/Fail Mode
:CONFigure:DEMod <char>
Title: Configure Demodulation Measurement
Description: Valid <char> parameter options: SUMMary|CDPower|CDPTable
This command configures the selected demodulation measurement. It
disables any other active measurements. Sets the instrument to single
sweep mode (:INITiate:CONTinuous OFF). Measurement settings can
be modified by using the [:SENSe]:DEMod commands before initiating a
measurement.
When the SUMMARY option is selected, the measurement results are
equivalent to the results that are displayed by choosing Measurements,
then Demodulator, then Modulation Summary from the front panel.
When the CDPower option is selected, the measurement results are
equivalent to the results that are displayed by choosing Measurements,
then Demodulator, then CDP from the front panel. When the CDPTable
option is selected, the measurement results are equivalent to the results
that are displayed by choosing Measurements, then Demodulator, then
CDP Table from the front panel.
Parameter: <char>
Front Panel
Access: Measurements, Demodulator
ООО "Техэнком" Контрольно-измерительные приборы и оборудование www.tehencom.com