4-13 :UNIT Subsystem Chapter 4 — Spectrum Analyzer Commands
4-48 PN: 10580-00341 Rev. C MS2720T PM
4-13 :UNIT Subsystem
The unit subsystem is used to modify the default units used for related parameters. These
changes affect parameters in both commands and responses.
:UNIT:POWer dBM|DBV|DBMV|DBUV|V|W
:UNIT:POWer?
Title: Measurement Units
Description: Sets the default amplitude units for input, output and display.
Available units: dBm, dBV, dBmV, dBµV, V, W.
Note that linear units are not operational with SPA V3.06. The set
command is non-operational with SPA V3.06 as well.
Parameter: DBM|DBV|DBMV|DBUV|V|W
Parameter Type: <char>
Default Value: dBm
Front Panel
Access: Amplitude, Units, Unit of Measure