Chapter 4 — Spectrum Analyzer Commands 4-7 :INITiate Subsystem
MS2720T PM PN: 10580-00341 Rev. C 4-31
4-7 :INITiate Subsystem
This subsystem controls the triggering of measurements.
:INITiate:CONTinuous OFF|ON|0|1
:INITiate:CONTinuous?
Title: Continuous/Single Sweep
Description: Specifies whether the sweep/measurement is triggered continuously. If
the value is set to ON or 1, another sweep/measurement is triggered
as soon as the current one completes. If continuous is set to OFF or 0,
the instrument enters the “idle” state and waits for the
:INITiate[:IMMediate] command or for :INITiate:CONTinuous
ON. The default value is ON. That is, sending :INIT:CONT is
equivalent to sending :INIT:CONT ON. The query version of the
command returns a 1 if the instrument is continuously
sweeping/measuring and returns a 0 if the instrument is in single
sweep/measurement mode. Note that rapid toggling between ON and
OFF is not allowed. The instrument must be allowed to make a full
sweep before toggling can be done.
Parameter: OFF|ON|0|1
Parameter Type: <boolean>
Default Value: ON
Related Command: :INITiate[:IMMediate]
Front Panel
Access: Shift-3 (Sweep), Sweep (Single/Continuous)