10-2 :CONFigure Subsystem Chapter 10 — GSM Commands
10-2 PN: 10580-00341 Rev. C MS2720T PM
10-2 :CONFigure Subsystem
This set of commands prepares the instrument for the selected measurement. It disables any
currently-enabled measurements and activates the specified measurement. It sets the
instrument to single sweep mode, waiting for an :INITiate command. It will not initiate the
taking of a measurement.
Current instrument settings may be changed to default values. These changes are identified
with their respective measurement commands.
:CONFigure:SUMMary
Title: Configure Summary Measurement
Description: This command configures the summary of all the related numerical
measurement results. It disables any other active measurements. None
of the instrument parameters are changed by the execution of this
command. Sets the instrument to single sweep mode
(:INITiate:CONTinuous OFF). Measurement settings can be modified
by using the [:SENSe] commands before initiating a measurement.
Front Panel
Access: Shift-4 (Measure), GSM/EDGE Summary
:CONFigure:DEMod
Title: Configure Demodulation Measurement
Description: This command configures the Demodulator measurement. It disables
any other active measurements. None of the instrument parameters are
changed by the execution of this command. Sets the instrument to
single sweep mode (:INITiate:CONTinuous OFF). Measurement
settings can be modified by using the [:SENSe] commands before
initiating a measurement.
Front Panel
Access: Shift-4 (Measure), Demodulator