14-3 :CONFigure Subsystem Chapter 14 — TD-LTE Commands
14-6 PN: 10580-00341 Rev. C MS2720T PM
14-3 :CONFigure Subsystem
This set of commands prepares the instrument for the selected measurement. It disables any
currently-enabled measurements and activates the specified measurement. It sets the
instrument to single sweep mode, waiting for an :INITiate command. It will not initiate the
taking of a measurement.
Current instrument settings may be changed to default values. These changes are identified
with their respective measurement commands.
:CONFigure SUMMary
Title: Set Measurement to TD-LTE Summary
Description: Sets the measurement to TD-LTE Summary. It disables any other
active measurements. None of the instrument parameters are changed
by the execution of this command. (:INITiate:CONTinuous OFF).
Measurement settings can be modified by using the [:SENSe]
commands before initiating a measurement.
Parameter: SUMMary
Parameter Type: <char>
Front Panel
Access: Measurements, TD-LTE Summary
:CONFigure?
Title: Configure Query
Description: :CONFigure? query returns the name of the measurement previously
set up using a CONFigure command or a MEASure? query. The list
below shows the possible return values.
SUMM
RF,SPEC
RF,ACLR
RF,SEM
RF, PVT, SUBF
RF, PVT, FRAM
RF,SUMM
DEM,CONS
DEM,CCPG
DEM,CCPT
DEM,PVRB
DEM,SUMM
PF
OTA,OTAS
OTA,OTATX
OTA,MAP