Chapter 4 — Spectrum Analyzer Commands 4-2 :CALCulate Subsystem
MS2720T PM PN: 10580-00341 Rev. C 4-13
:CALCulate:LIMit:VERTical:ADD
Title: Add Vertical
Description: Adds vertical. This will add two inflection points that share the same
frequency and are centered midpoint between adjacent points. Use
:CALCulate:LIMit:TYPe to set the currently active limit line.
Related Command: :CALCulate:LIMit:TYPe
Front Panel
Access: Shift-6 (Limit), Limit Edit, Add Vertical
:CALCulate:LIMit:X <x-parameter>
Title: Limit X Value
Description: Sets the location of the active limit line on the x-axis at the specified
location. This moves the entire limit and moves the active limit point to
the given value. <x-parameter> is defined in the current x-axis. Note
that this will change the Move Limit on the front panel to Limit if it is
currently set to Point. Use :CALCulate:LIMit:TYPe to set the currently
active limit line.
Parameter: <x-parameter>
Default Unit: Hz or for zero span in seconds
Example: To move the limit and set active point to 5 Hz:
:CALCulate:LIMit:X 5
:CALCulate:LIMit:X 5Hz
To move the limit and set active point to 500 MHz:
:CALCulate:LIMit:X 500MHz
To move the limit and set active point to 2.5 seconds (In zero span):
:CALCulate:LIMit:X 2.5
:CALCulate:LIMit:X 2.5s
To move the limit and set active point to 25 µs (In zero span):
:CALCulate:LIMit:X 25µs
Front Panel
Access: Shift-6 (Limit), Limit Edit, Frequency