EasyManuals Logo

APT 320XAC User Manual

APT 320XAC
161 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #143 background imageLoading...
Page #143 background image
137
key at this screen you return to the calibration mode screen.
Enter the current reading from the ammeter with the numeric keypad. When the
value has been selected press the Enter soft key and you will be moved to the
next calibration parameter Volt 150.0V. The soft keys Volt+ and Volt- are
available, if needed, to adjust the voltage output of the instrument to set a proper
peak current value if a non-recommended load is used. If you press the Esc soft
key you will be returned to the calibration mode screen.

Table of Contents

Other manuals for APT 320XAC

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the APT 320XAC and is the answer not in the manual?

APT 320XAC Specifications

General IconGeneral
BrandAPT
Model320XAC
CategoryAC Power Distribution
LanguageEnglish

Related product manuals