Do you have a question about the APT 300XAC Series and is the answer not in the manual?
Input Frequency | 50/60 Hz |
---|---|
Input Current | 30A |
Output Frequency | 50/60 Hz |
Maximum Current per Phase | 30A |
Mounting | Rackmount |
Operating Temperature | 0°C |
Humidity | 95% non-condensing |
Dimensions | 19 x 1.75 inches |
Certifies instrument meets specifications. Details warranty period, repair process, and conditions that void warranty.
Presents detailed electrical specifications for AC and DC output, including voltage, current, power, frequency, and measurement accuracy.
Describes system parameters that change overall instrument operation, applying universal changes across memories and steps.
Guides on editing system parameters like Auto Run, Out Mode, Alarm, and Contrast for instrument configuration.
Explains how to use memories and steps for programming test routines, including selecting and naming memories.
Describes various test parameters that can be edited for each memory and step location, not universally applied.
Details how to edit specific test parameters like Start Angle, End Angle, Voltage, Frequency, and Transient functions.
Guides the user on initializing a test in PROGRAM Mode, including AC and DC output procedures.
Guides the user on initializing a test in MANUAL Mode, including AC and DC output procedures.
Details how to configure instruments for Parallel and Polyphase modes using the Option 08 Linking Card.
Guides on initializing tests in Parallel or Polyphase modes, noting slave unit displays are not editable.
Provides a comprehensive list of commands for controlling the instrument and retrieving data via USB, GPIB, and RS-232 interfaces.
Details the hardware calibration procedure, including required standards and equipment, for verification before software calibration.
Guides on connecting a DVM and adjusting VR1 to set the control circuit power voltage to +15V +/- 0.1V.
Guides on connecting a DVM and adjusting VR1 to set the amplifier inverter DC bus voltage to +320V +/- 0.3V.
Details connecting DVM, using jumper JP1, and adjusting VR4/VR2 to set DC offset voltages for 0V and 115V.
Guides on setting output voltage, noting wattmeter gain differences, and adjusting VR1 to match readings.
Explains connecting load resistors, setting output voltage, and adjusting VR1 until OCP is displayed on the meter.
Recommends software calibration after hardware verification, listing required measurement standards.