PRODUCT#.
The Autostart feature will not work with the SERIAL# setting.
The Autostart feature will enable the instrument’s output once the product
number barcode is scanned a second time when in the PRODUCT# and
SER/PROD modes. Do not touch the DUT at any time when using this feature in
order to avoid potential shock or serious injury.
Use the Change soft key to select the Autostart setting. Press the Enter soft key
to accept the new setting or the Exit soft key to cancel and return to the original
setting.
11.4.13 Ethernet Card Settings Commands and Companion Queries
1=Manual, 0=Auto
(DHCP/BOOTP)
Dotted decimal form. Ex.
192.168.1.50
8 character max, must start with
a letter
Example response:
00:20:4A:8B:B4:30
0=Off, 1=Serial# and Product#,
2=Serial# Only, 3=Product#
Only, 4=Run File
Communication Considerations
All of the above commands (excluding the query commands) will respond
with the 06 hex (6 decimal) Acknowledge (ACK) ASCII control code if the
transfer was recognized by the instrument.
If there was an error with the command string, the instrument will respond
with 15 hex (21 decimal), the Not Acknowledge (NAK) ASCII control code.