136
Follow the prompt message provided on the display, and press the Test/Reset
button to move into the calibration screen for wattage. If you press the Exit soft
key at this screen you return to the calibration mode screen.
Enter the power reading from the wattmeter with the numeric keypad. When the
value has been selected press the Enter soft key and you will be moved to the
next calibration parameter. The soft keys Volt+ and Volt- are available, if needed,
to adjust the voltage output of the instrument to set a proper wattage value if a
non-recommended load is used. If you press the Esc soft key you will be
returned to the calibration mode screen.
10.16 Calibration of Peak Current
Use the , soft keys to navigate to the A-Peak xx.xA parameter and press the
Select soft key.
Follow the prompt message provided on the display, and press the Test/Reset
button to move into the calibration screen for current. If you press the Exit soft