To calculate exactly the dosing, the following formulas can be applied:
4.10 CALIBRATION OF PH ELECTRODE
The calibration of the pH electrode is carried out on two points, therefore it is requested to have the following
material always available:
pH 7 buffer solution.
pH 4,01 buffer solution.
Container with water for cleaning the electrode (i.e. a glass).
Select the option pH Calibration from the installer menu and follow the instructions displayed.
Once carried out the pH 7 calibration, rinse the electrode with water, so to prevent the pH 4 solution from
polluting.
Once the procedure is completed, the quality of the calibrated electrode will be displayed; if the quality of the probe
results 25% or less, carry out a new calibration.
Should the result still be 25% or less, it is recommended to replace the probe.
4.11 CALIBRATING THE REDOX ELECTRODE
(it is displayed for Time/pH/Rx and pH/Rx technopool3)
The calibration of the redox electrode is carried out only on one point, therefore it is requested to have the following
material always available:
475 mV buffer solution.