14. Unit Set Up
14.1 . Load Cell Calibration procedures
Step 1 – Tare (Calibrating the zero weight display 0.00kg with the empty cylinder on platform)
Place empty cylinder or similar weigh (as unit
empty cylinder) on the cylinder platform.
Connect the cylinder hoses to the cylinder.
Check that cylinder and the platform are free
(not touching the body of the machine) and
the strap is properly secured.
Press and hold depressed the "UP" button
and switch ON the unit (on the main power
switch at back).
The display will indicate TARE.
Now, release 'UP' button.
After BEEP sound, press 'START' button to
confirm the TARE. The display will indicate
“DONE”.
Note: If “DONE” is not indicated, perform the
function again.
The display will indicate 0.00kg which is the
right weight because the cylinder is empty.