8. Troubleshooting and Alarm Conditions
The following table provides a troubleshooting and alarm condition guide for the Alpha Active 3
system in the event of malfunction. These alarms do not cause any delay or interruption in
therapy.
Indicator Possible Cause Remedy Priority
LOW PRESSURE The tube-set is not
connected properly.
CPR not fully closed.
There is a leak in the
system.
Check the tube-set
connector and ensure
it is securely fitted to
the pump.
Close CPR unit.
Call service engineer.
Low priority
According to
IEC60601-1-8
POWER FAILURE Power has been
removed from the
pump.
Re-apply power or
switch the pump off.
Low priority
According to
IEC60601-1-8
SERVICE Pump has detected an
internal fault, such as
the gearbox failure.
Switch the pump off
and call service
engineer.
Low priority
According to
IEC60601-1-8