Electromagnetic Compatibility
Electromagnetic Compliance
The Voyager has been tested for compliance with current regulatory standards regarding its capacity to
block EMI (electromagnetic interference) from external sources.
Nonetheless, some procedures can help reduce electromagnetic interferences:
Ensure that other devices in patient-monitoring and/or life-support areas comply to accepted emissions
standards.
Maximize the distance between electro-medical devices. High-powered devices may produce EMI that
can affect the ceiling lift.
For more information on how to manage the unit’s RF electromagnetic environment, please consult the
AMI TIR 18-1997 - Guidance on Electromagnetic Compatibility of Medical Devices for Clinical/Biomedical
Engineers.
WARNING: 8VHRIDFFHVVRULHVFDEOHVDQGVSDUHSDUWVRWKHUWKDQWKRVHVSHFL¿HGRUSURYLGHG
by Arjo could result in increased electromagnetic emissions or decreased electromagnetic
immunity of this equipment and result in improper operation.
WARNING: Use of this equipment adjacent to or stacked with other equipment should be avoided
because it could result in improper operation. If such use is necessary, this equipment and the
other equipment should be observed to verify that they are operating normally.
WARNING: The equipment may cause radio interference or may disrupt the operation of nearby
equipment. It may be necessary to take action, such as reorienting, relocating the equipment or
shielding the location.
Electromagnetic Emissions
Guidance and Manufacturers Declaration -
Electromagnetic Emissions - For all Equipment and Systems
The Voyager is intended for use in the electromagnetic environment indicated below. The customer or
the user of the Voyager should assure that it is used in such an environment.
Emissions test Compliance Electromagnetic environment - guidance
RF emissions
CISPR 11
Group 1
The Voyager uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely
to cause any interference in nearby electronic equipment.
RF emissions
CISPR 11
Class B
The Voyager is suitable for use in all establishments,
including domestic establishments and those directly
connected to the public low-voltage power supply network
that supplies buildings used for domestic purposes.
Harmonic
emissions
IEC 61000-3-2
Class A
Voltage
ÀXFWXDWLRQVÀLFNHU
emissions
IEC 61000-3-3
Complies