Cleaning, disinfection and sterilization
1590 99 002_1 - 21. February 2011 69
INSTRUCTIONS
Cleaning:
manual
Make sure that
these cleaning agents only get in contact
with glass or metal surfaces of the handpiece.
The opt
ical surface must not get scratches and afterwards it
must be lint-free and free of streaks.
Cleaning:
automatic
not possible
Disinfection
The accessible outer surfaces are to be disinfected by means of a
disinfection cloth (e.g. Kodan cloth)
. When doing this follow the
.
Sterilization not possible, not required
Package for sterilized parts not applicable
Storage and transport The parts must be stored in a clean condition and protected
against contamination. Ensure that the parameters of the
correspondingly admissible ambient conditions indicated in
chapter Technical Data are given.
The manufacturer has validated that the above instructions are SUITED for preparing the accessory
parts for their reuse.
WARNING
The contact
There is the risk of microorganisms being transferred.
The person who cares about cleaning has all the responsibility that the actually performed
cleaning and disinfecting works by means of the equipment and materials and with the
staff involved achieve the desired results.
To guarantee this, validations and routine monitoring of the procedures are normally required. The
effect and adverse consequences of each deviation from the instructions given should also be
evaluated.