76 1590 99 002_1 - 21. February 2011
11.2 Regular safety check
SCOPE OF ANNUAL SAFETY CHECK
The safety check should be carried out in compliance with the relevant national legally
regulations, as amended, and include at least the tests mentioned below. The tests should be
run in the order specified:
visual inspection of laser device and accessories
measurement of protective ground wire resistance
earth leakage test under normal conditions
measurement of the actual output power of the laser modules at the handpiece; calibration
of laser power, if necessary
functional test.
Tip
To inform yourself about the extent and performance of the safety checks, please
contact your distributor or directly Asclepion Laser Technologies GmbH.
Additional information
Upon request, the manufacturer will provide circuit diagrams, component parts lists, descriptions,
calibration instructions, or other information not yet contained in this User Manual to assist
appropriately qualified technical staff to repair those parts of the device that have been designated
his device and that was authorized
to repair it.