EasyManua.ls Logo

ATAGO PR-32a - Performing Zero Setting

ATAGO PR-32a
28 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
9
5.
Zero setting
(In case of PR-32
α
, PR-101
α
, PR-201
α
)
CAUTION
◇ The refractometer needs the "zero setting" (calibration) before it is used for the first
time in a day. Moreover, if the ambient temperature changes during the daily work with
the refractometer, it needs the zero setting again.
◇ The zero setting is also needed whenever the battery is replaced.
◇ Do not use any metallic implement for dropping water on the prism because the metal
tool possibly damages the prism surface.
3.5℃
0 0
0
3.5℃
[ 1
]
Flashes 3 times
Measurement
value is d
isplayed
Zero-set correctly.
Completely cover prism surface
Min. 0.1mL
Clean the prism
surface thoroughly.
Apply tap water.
Press the
START/OFF Button.
Measurement:not 0.0%
Measurement:0.0%
Press the
ZERO Setting
Button.
Zero setting is complete.
Flashes
3 times
000 is
displayed.
Perform zero-setting
Confirm zero-setting
20.4℃
0
0
.
0
20.4℃
0
0
.
0
20.4℃
0
0
.3
3.5℃
0 0
0