11
7.
Measurement
WARNING
When measuring a substance harmful to the human body, be well aware of its properties
and put on protective gloves, mask, etc.
CAUTION
Do not use any metallic implement for sampling because the metal tool possibly damages
the prism surface.
Do not apply water or sample over any part other than the surface of the prism.
Application on any other part of the instrument may result in a malfunction.
N Refer to "12.Methods of measurement for special
samples"(Cp.18) when measuring special
samples (such as hot or high viscosity samples).
Measurement value is displayed
Example:PR-101α
Clean the prism
surface thoroughly.
Press the
START/OFF Button.
3.5
[ 1
]
Apply a sample.
Flashes 3 times
Completely cover prism surface
Min. 0.1mL
20.4