D620.388.00_QS_UK_01
38 / 42
PARAMETERS menu
Use this menu to configure the measurement cycle associated to each test program.
TYPE : DIRECT FLOW
COUPL A 0.0 s
0.0 s
0.0 s
Inf. s
0.0 s
bar
0.000
FILL TIME
STAB TIME
TEST TIME
DUMP TIME
Press. UNIT
Max FILL
PARAM / Pr001
Default parameters of the FLOW type tests
Label Parameter Description
COUPL. A or COUPL. B
Coupling time Required times when instrument manage
automatic jigs
FILL TIME
Fill time Time to pressurise the part under test
STAB TIME
Stabilization time Time to stabilise the flow
TEST TIME
Test time The flow in the part must be between the
minimum and maximum reject level.
Press. UNIT
Pressure units Pressure unit (bar, mbar, PSI, Pa, kPa, MPa)
Max PRESS
Maximum fill pressure Maximum level of the fill pressure
Min PRESS
Minimum fill pressure Minimum level of the fill pressure
FLOW UNIT
Reject unit Flow rate unit displayed:
SI System: ml/h, l/h, ml/s, ml/min, sccm, l/min,
m3/h
US System: in3/h, ft3/h, in3/min, in3/s, sccm.
Max Flow
Maximum reject Upper flow rate limit. Above this limit, the part is
considered as defective.
Min Flow
Minimum reject Lower flow rate limit. Under this limit, the part is
considered as defective.
FUNCTIONS
Functions Access to additional functions
Additional functions
Label Function Description
24V OUTPUTS
Auxiliaries output 24 V Available outputs for external automatism
ANALOG OUTPUT
Analog output Duplicate the flow and pressure measurements
on the analogues outputs
ATR1 / ATR2
ATR 1 - 2 Specific filters on flow measurement
AUTO CONNECT
Automatic connector Function to manage automatic jigs
AUTO MODE
Automatic mode Starts automatically when pressure reach min
pressure level
BUZZER
Buzzer Buzzer activation configuration
CODE READER
Bar code reader Bar code configuration
DISPLAY MODE
Display Mode Flow measurement resolution
END OF CYCLE
End of cycle Several automatism case depending on fail part
management
FILL MODE
Fill types Special filling methods