9.0 Technical data
Voltage 220-240V~±10%;50/60Hz
Special voltage 100-127V~±10%;50/60Hz
Current consumption max.6.8A(220-240V~);14.8A(100-127V~)
Powerconsumption max. 1500 W
Connections Powerconnection;controloutputforanystagmograph;equipotential
equalization;watersupplyG3/4"externalthread;waterdrainageG1/4"
external thread; connection for double-barrelled hose
Fuses F8A(f.220-240V~);F15A(f.110-127V~)
Stimulation time Adjustable by timer from 1 up to 99 sec.
Timer indication Indication accuracy ± 0.5 s ± ½ digit
Water temperature 20°C - 47°C
Temperature indication Indicationaccuracy±0.6°C±½digit(canonlybeguaranteewith
unchangedambientconditions)
Feedtemperatureofthewater +7...+19.5 °C
Water quality Drinkingwater
Waterowintheirrigationmode
(highow)
450 ml/min: Standard unit and variant 250/450 ml/min
400 ml/min: Variant 400/400 ml/min
500 ml/min: Variant 500/500 ml/min
Waterowinthestimulationmode
(lowow)
150ml/min:Standardunit150/450ml/min(adjustable)
250ml/min:Variant250/400ml/min(adjustable)
400ml/min:Variant400/400ml/min(adjustable)
500ml/min:Variant500/500ml/min(adjustable)
Operating time Continuous operation
Operating pressure Atleast:+2000hPa(2bar);max:+6000hPa(6bar)
Protective earth conductor resistance
Earth leakage current
Enclosure leakage current
Patient leakage current
max.0,1Ω
max. 0.5 mA
max. 0.1 mA
max. 0.1 mA
Ambient conditions
Transport / storage
-20...+50°C
5...90%airhumiditywithoutcondensation
air pressure of 700...1060 hPa
Ambient conditions
Operation
+10...+35°C
20...80%airhumiditywithoutcondensation
air pressure of 700...1060 hPa
Maximumoperationalaltitude ≤3000m(NN)
Contamination level Class 2
Overvoltage category II
DimensionsHxWxD 14.5 x 37 x 31.5 cm
Weight Approx.5,5kg
Period tests Inspectionaccordingtothemanufacturersspecicationsevery12months.
Safetyclass(EN60601-1) I
Degree of protection Type B
Protection class IPX0
Furtherclassicationsaccordingtoother
regulations
VDEprotectionclass1(IEC601/EN60601)
ClassicationaccordingtoAppendixIXEC
Directive 93/42/EEC
Class IIa
CE marking CE 0124