11.0 Notes on EMC
11.1 Guidelines and Manufacturer´s Declaration - Emissions
TheATMOS
®
Variothermplusisdesignedforoperationintheenvironmentspeciedbelow.ThecustomeroruseroftheATMOS
Variotherm plus should ensure that it is used in such an environment.
Emissions Test Compliance Electromagnetic Environment - Guidance
RF Emissions acc.to CISPR 11 Group 1 TheATMOSVariothermplususesRFenergyonlyforits
internalfunction.Therefore,itsRFemissionsareverylowand
are not likely to cause any interference in nearby electronic
equipment.
RF Emissions according to CISPR
11
Class B
TheATMOSVariothermplusissuitableforuseinall
establishments,includingdomesticandthoseconnected
directlytoapublicpowersupplynetworkthatsupplies
buildings used for residential purposes.
Harmonicemissionsaccordingto
IEC 61000-3-2
Class B
Voltageuctuations/ickeraccording
to IEC 61000-3-3
Corresponds
11.2 Guidelines and Manufacturer's Declaration - Immunity
TheATMOSVariothermplusisdesignedforoperationintheelectromagneticenvironmentspeciedbelow.Thecustomeror
useroftheATMOSVariothermplusshouldensurethatitisusedinsuchanenvironment.
Immunity Test IEC 60601- Test Level Compliance Level
Electromagnetic Environment -
Guidance
Electrostatic discharge
(ESD)accordingtoIEC
61000-4-2
± 6 kV Contact
± 8 kV Air
± 6 kV Contact
± 8 kV Air
Floorsshouldbemadeofwoodor
concreteortiledwithceramictiles.If
oorsaresynthetic,therelativehumidity
shouldbeatleast30%.
EFT IEC 61000-4-4 ±2kVMains
± 1 kV I/Os
±2kVMains
Inapplicable
Mainspowerqualityshouldbethat
of a typical commercial or hospital
environment.
Surges IEC 61000-4-5 1 kV
Differential
1 kV
Common
2 kV
Differential
1 kV
Common
Mainspowerqualityshouldbethat
of a typical commercial or hospital
environment.
Magneticeldatpower
frequency50/60Hzacc.
to IEC 61000-4-8
3 A/m Inapplicable Powerfrequencymagneticeldsshould
be that of a typical commercial or
hospital environment.
MedicalelectricalequipmentissubjecttospecialprecautionswithregardtoEMCandmustbeinstalledacc.tofollowing
EMCnotes.
PortableandmobileHFcommunicationfacilitiescaninuencemedicalelectricalequipment.
Theuseofotheraccessories,otherconvertersandcablesthanstatedmayleadtoanincreasedemissionorareduced
interference immunity of the equipment or system.
i
Thedevicemaynotbeuseddirectlynexttootherdevicesorpiledupwithotherdevices.Ifoperationnexttoorpiledwithother
devicesisnecessary,pleasewatchthedevicetocheckitsintendedoperationinthisarrangement.
i