2.0 For your safety
2.1 Important instructions for the maintenance of the hygiene status of warm water
irrigation units
Forlooseningcerumenintheauditorymeatusandforthestimulationofthelabyrinthinesystem,ATMOSoffersthewarmwater
unitsHygrothermplus(37°C)andVariothermplus(20°-47°C).
Theseunitsheatthedrinkingwaterwhichcomesfromthehouseholdconnectiontothepreselectedtemperature.Theoperating
instructionsmustbefollowed,ortheincreasednumberofpathogensintherinsewatercanaffectthehealthofsensitive
patients.
Requirement for taking into use
ThewaterwhichisprovidedbythehouseholdconnectionmustatleastmeettheWHOguidelinesorthecountry-specic
guidelinesfordrinkingwater.
Connection
Freshwatermustowthroughthewallconnectionforaperiodofapprox.1minutepriortotheunitbeingconnected.*
Ensurethattheinstallationishygienicallycarriedout!Disinfectthethreadedconnectionswith70%alcoholbeforescrewing
themintoplace.*
Weardisposablegloveswhencarryingoutmaintenanceworkonpartswhichcomeintocontactwiththewater.*
Carryoutthermalcleaningatleast3timesbeforepatientuse.*
*ThisworkiscarriedoutbytheATMOSservicetechnician.
Running operation
Beforebeginningthesurgicaloperation,switchontheunitandawaitthethermalcleaningprocess.
Afterthedevicehadbeenidleforalongperiodoftime,(weekend,duringholidays,etc.)thethermalcleaningprocessmust
be initiated and repeated.
Werecommendthatthedevicebeswitchedoffandonevery2hoursandnotlongerthan4hourstostartthethermal
cleaning mode.
Exchange and disinfection required for contact parts such as the jet connection after each patient (prevention of retrograde
germication).
Beforeeveryuse,spraywaterandcheckthetemperature.
Service
Ensurethatthesuspendedparticlelterishygienicallychanged(seecommissioning)(disposeoflterandcarefullycleanthe
lterglass).Ensurethatnocontactismadewiththecontaminatedparts.
Carry out thermal cleaning after each service.
Werecommendthatthenumberofgermsinthewaterberegularlydeterminedattheinputandoutputoftheunit.Inthecase
ofacomplaint,datameasurementsmustbeprovewhetherthecontaminantsare"water"or"skin”bacteria.