Maintenance
AVL 415SE Smoke Meter — Maintenance Guide
2.2.2
Service Engineer - Maintenance Schedule
For the disposal of environmentally sensitive substances, see the AVL 415S
Smoke Meter Operating Manual, chapter Disposal of Environmentally Sensitive
Substances.
Clamping lever Annually Visual check. If damaged or if any part
is bent off.
Pump Annually Check and replace if necesary. Every 2 000 - 3 000 oper-
ating hours.
Nipple & hose connec-
tions
Annually Replace. Annually
Cleaning of electronical
boards
Annually Clean with high purity compressed
air.
If necessary or defective.
Filters of shop air filter
unit
Annually Check and replace if necessary. once every 1 … 2 years;
much more often for oil
lubricant compressors:
at least: every
1…2months
Check valve Annually Check and replace if necesary. Annually
Pressure reduction
assembly
Annually Check. After assembly.
Pressure switch of shop
air option
Annually Check and adjust if necessary. Annually
Component / Test Lifetime /
Service interval
Maintenance Replacement
Tab. 1
WARNING
Fire hazard
When using the required appliances like sprays, mounting pastes, lubricants,
always observe the relevant handling instructions and safety measures of the
various products.
All sprays used are a fire hazard and highly inflammable. Installation work using
the appliances on and inside measurement devices and on test beds must only
be carried out on cooled-down parts and in well ventilated areas.
NOTICE
The compressed air used to clean, has to be absolutely free from oil, water, and
particles.