MagneW
3000 Electromagnetic
Flowmeter
Hyper-Fill
Model: MGR13C (Converter)
MGR11U (Detector)
Users Manual
CM2-MGR100-2001
2nd edition
Question and Answer IconNeed help?

Do you have a question about the Azbil MagneW 3000 and is the answer not in the manual?

Azbil MagneW 3000 Specifications

General IconGeneral
BrandAzbil
ModelMagneW 3000
CategoryMeasuring Instruments
LanguageEnglish

Summary

Product Overview

Instrument Installation Guide

Converter Installation Procedures

Mounting the converter using M4 screws and handling the rubber bush for isolation.

Detector Installation Guidelines

Installing the detector in the environment, connecting cables, and ensuring proper pipe alignment.

Wiring Connections

Details on connecting converter wiring, pulse output wiring, and master-slave wiring.

Converter Configuration Settings

Setting the Flow Range

Selecting an appropriate range for the flowmeter, aiming for approximately 70% flow rate.

Pulse Scale Configuration

Choosing between 0.1 mL/P or 0.05 mL/P pulse scales for flow measurement.

Pulse Width Settings

Selecting pulse width options of 30 µS or 200 µS and their tolerances.

Drop Out Adjustment

Setting the dropout value (0-10%) to prevent incorrect totalization due to flow pulsation.

Master/Slave Configuration

Configuring converters in master/slave mode to minimize magnetic interference in multi-head applications.

Operational Procedures

Warm-up and Stabilization

Ensuring 30-minute warm-up for electronics stabilization and removing bubbles for accurate measurement.

Zero Adjustment Process

Performing zero adjustment using a variable resistor or PLS ZERO button for flow rate accuracy.

Span Adjustment Procedure

Tuning the span using the 'SPAN' variable resistor to correct discrepancies between actual and measured values.