B. Braun Vista basic - Additional Functions with Inserted Service Plug; Disabling the Occlusion Sensor

B. Braun Vista basic
55 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Service
Program
Additional
Functions
with
inserted
Service
Plug
Software
Version
andUser
Languages
1.
Insert
service
plug
into
the
MFC
socket
at
the
rear
of
the
unit.
2.
Switch
on
unit
and
keep
the
ON/OFF
button
pressed
(for
max.
15
s).
3.
The
software
version
and
user
languages
are
displayed
inthe
LCD.
4.
The
unit
is
switched
on
when
the
ON/OFF
button
is
released.
5.
**
appears
in
the
LCD
if
the
service plug
is
inserted,
Two
LEDs
are
integrated
in
the
service
plug:
-
green
=
Power
supply
active
-
red
=Alarm
The
following
conditions
are
active:
-
The
operating
alarms
are
muted.
-
All
special
functions
are
accessible
(including
the
disabled).
-
The
special
functions
are
slightly
modified.
(Example:
SM
battery
capacity
has
key
for
0
min/32
min
presetting).
-
The
battery
capacity
display
switches
between
the
nominal
i
)
and
the
actual
capacity.
If
the
maximum
battery
capacity
is
i
not
reached
”?”
will
be
displayed
before
the
new
nominal
ca-
pacity.
-
The
tube
type
will
be
displayed
in
the
tube
selection
menu.
Disabling
the
Occlusion
Sensor
The
electronic
pressure
monitoring
can
be
deactivated
to
check
the
mechanical
occlusion
pressure.
1.
Select
“Occlusion
Pressure”
with
the
SM
key.
2.
Then
select
“mechanical”.
3.
Return
to
the
main
menu
with
END.
A
too
low
pump
speed
is
indicated
in
the
display
by
"Pressure
Alarm”
with
underlined
stars.
The
display
"Pressure
Alarm”
with
underlined
questions
marks
indicates
that
no
tube
has
been
inserted.
Vista
basic
950971
Rev
A
(5/04)