Unit
Elements
-
Delivery
rate
min/max.
-
Air
alarm:
air
bubbles
in
ml
and
air
rate
in
mi/h
respectively
-
User
language
-
Special
functions
(ON/OFF)
-
Menu
-
Staff
call
type
If
data
is
not
entered,
“Calibration
Defective”
may
be
displayed
after
the
unit
is
switched
on
again.
9.
After
ending
the
service
program
save
the data.
The
counters
for
operating
hours,
battery
operation
and
pump
head
cycles
are
reset
to
zero
when
the
pump
unit
is
exchanged.
If
“Calibration
Defective’
is
displayed,
select
the service
program
and
quit
with
“Save?
Yes”.
Check
Safety check,
pump
unit
check.
Vista
basic
950971
Rev
A
(5/04)
4-13