Model M-2000 Mag Meter
Advanced
Analog Calibrate Custom Settings
[Zero Scale: 0 mA]
[Full Scale: 0 mA]
S
Scale/Full Scale offsets on site.
To set the analog calibration custom settings, follow these steps from the
Advanced menu:
1. Select Analog Calibrate to view the Analog Calibrate menu.
2. Select Custom Settings to view the Custom Settings display.
3. Select one of the following:
• Offset 4 mA
• Offset 20 mA
4. Configure desired offset.
5. Press [E] to save the option and to return to the Custom Settings menu.
6. Press [E] to return to the Analog Calibrate menu.
Factory Settings
[Factory Set]
S
To change the analog calibration factory settings, follow these steps from the
Advanced menu:
1. Select Analog Calibrate to view the Analog Calibrate menu.
2. Select Factory Settings to view the Factory Settings display.
3. Select one of the following:
• Calibration Point A
• Calibration Point B
4. Set the calibration point to the measured output current.
5. Press [E] to save the option and to return to the Custom Settings menu.
6. Press [E] to return to the Analog Calibrate menu.