4-11
Installation & Operation Manual
Info/Help
Error Counts
[0]
U
Description This menu provides a diagnostic view of the meters performance. Below are several
system diagnostic counters and their definitions. Use discretion when interpreting
these counters. These values could be altered during system setup or when using
the verification device. We suggest that you reset these counters before you start
monitoring your system and look for conditions possibly affecting performance.
Detector The number of times an invalid detector condition has been observed.
Empty Pipe The number of times an empty pipe condition has been observed by the meter.
Full Scale The number of times the flow has exceeded the full scale flow setting.
Totalizer The number of times the totalizers have exceeded limits of the meter.
Pulse Sync. The number of times the pulse outputs have fallen out of synchronization.
ADC Interrupt The number of times an analog input measurement has been missed.
ADC Range The number of times the analog input measurement range has been exceeded.
System Error A diagnostic system message indicating the reason for a system reset.
System Resets The number of times the meter has been reset.
System Reset ID Diagnostic information about a system reset as a result of expired internal timers.

Table of Contents

Other manuals for Badger Meter M Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Badger Meter M Series and is the answer not in the manual?

Badger Meter M Series Specifications

General IconGeneral
BrandBadger Meter
ModelM Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals