Model M-2000 Mag Meter
Menu Manager Conguration Errors
Error Description Recommended Action
110 Output 1/2: Pulse Output
Conguration Error
This error is observed when improperly conguring the full scale ow, pulse per unit, or pulse width.
This error can indicate the following conguration violations:
1. Pulse frequency exceeds limits at full scale ow
2. Pulse duty cycle is less than 50% at full scale ow (pulse on time > pulse o time)
Pulse frequency limit is 10 kHz. However, with a non-zero pulse width conguration, the limit is 500
Hz to achieve a 50% pulse duty cycle.
If not using the pulse outputs, set the pulses per unit to zero to allow for reconguration of the full
scale ow. If it is required to use the pulse outputs, reevaluate the pulse output conguration.
Consider recording and clearing totalizers prior to changing totalizer units.
120 Display: Totalizer Conversion
Error – Totalizer cannot be properly
converted for display
This error is observed while trying to change the totalizer units. Limits of display will prevent
improper conguration of the volume unit dependent on current totalizer values. Consider
recording and cleaning totalizers prior to changing totalizer.
121 Output 1/2: Pulse Output
Conguration Error
This error is observed when changing the totalizer units of measure. This error implies the pulse
conguration exceeds limits (see error 110). Please note the pulses per unit is not automatically
updated on volume unit reconguration. The pulses per unit should be manually changed to
accommodate the desired units of measure. It may be necessary to set the pulses per unit to zero
then change to the desired totalizer units.
140 Output 3: Conguration Error – Full
scale frequency exceeds limits of
relay (1000 Hz)
Reduce full scale frequency setting of output when hardware is congured for relay operation.
150 Output 3: Conguration Error – Full
scale frequency exceeds limits (10
kHz)
Reduce full scale frequency setting of output when hardware is congured for open collector
operation.
Display Error / Status Messages
Error Message Possible Cause Recommended Action
Err: Detector No detector connection with amplier. Check detector and cable connections in accordance with this manual.
Connection between amplier and detector Contact Technical Support.
Supply voltage too low. Contact Technical Support.
Grounded coils in meter. Contact Technical Support.
Water in detector. Contact Technical Support.
Err: Empty pipe Pipe may not be full. Make sure all trapped air is out of system.
If uid or uid conductivity recalibrate the parameter.
Err: Full scale Actual ow rate is exceeding programmed ow. Reduce ow rate or increase the programmed full scale value by more than 5%.
Err: AD-Range AD-Converter is exceeding signal limits. Check the grounding scheme of the meter installation. See grounding section in
this manual. Verify pipe is not empty.
Err: AD-INT Initialization of AD-Converter unsuccessful. Contact Technical Support.
Err: Rollover Rollover counters have exceeded limit. Clear all totalizers.
Err: Rollover Status Totalizer rollover has occurred. Reload totalizer then clear all totalizers.
Err: Simulation I/O simulator is enabled. Disable simulator in I/O menu.

Table of Contents

Other manuals for Badger Meter M Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Badger Meter M Series and is the answer not in the manual?

Badger Meter M Series Specifications

General IconGeneral
BrandBadger Meter
ModelM Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals