7. Important information
7.4 EMC notice
General information
No specic requirement on the use of external cables or other accessories except power supply.
With the installation of the device, use only the delivered power supply or a spare part provided by the
legal manufacturer. Using another can result in a decrease of the immunity level of the device.
Electromagnetic emissions
The Eonis is intended for use in the electromagnetic environment specied below. The customer or the
user of the Eonis should assure that it is used in such an environment.
Emissions test Compliance Electromagnetic environment
Guidance
RF emissions
CISPR 11
Group 1 The Eonis uses RF energy only
for its internal function. Therefore,
its RF emissions are very low
and are not likely to cause any
interference in nearby electronic
equipment.
RF emissions
CISPR 11
Class B
Harmonic emissions
IEC 61000-3-2
Class D
Voltage uctuations/ icker
emissions
IEC 61000-3-3
Complies
The Eonis is suitable for use
in all establishments, including
domestic establishments and
those directly connected to the
public low-voltage power supply
network that supplies buildings
used for domestic purposes.
This Eonis complies with appropriate medical EMC standards on emissions to, and interference from sur-
rounding equipment. Operation is subject to the following two conditions: (1) this device may not cause
harmful interference, and (2) this device must accept any interference received, including interference that
may cause undesired operation.
Interference can be determined by turning the equipment off and on.
If this equipment does cause harmful interference to, or suffer from harmful interference of, surrounding
equipment, the user is encouraged to try to correct the interference by one or more of the following mea-
sures:
Reorient or relocate the receiving antenna or equipment.
Increase the separation between the equipment and receiver.
Connect the equipment into an outlet on a circuit different from that to which the receiver is connected.
Consult the dealer or an experienced technician for help.
Electromagnetic immunity
The Eonis is intended for use in the electromagnetic environment specied below. The customer or the
user of the Eonis should assure that it is used in such an environment.
Immunity test
IEC 60601
Test levels
Compliance level
Electromagnetic
environment
guidance
Electrostatic discharge
(ESD)
IEC 61000-4-2
± 6kV contact
±8kVair
± 6kV contact
±8kVair
Floors should be wood,
concrete or ceramic tile.
If oors are covered with
synthetic material, the
relative humidity should
be at least 30%
28 K5903081 EONIS 08/03/2016