4. Maintenance
4. MAINTENANCE
4.1 Scheduled maintenan ce
About
The Nio Color 2MP 23" does not require any scheduled maintenance or calibration activities. We recom-
mend to use QAWeb with the Barco default tests and frequencies to calibrate and maintain the display,
or to return the display to a Barco approved maintenance organization. In any case of doubts, please
contact Barco Healthcare.
4.2 Clea ning
WARNING: Unplug the power cable from the mains power input before cleaning the
display.
CAUTION: Take care not to da mage or scratch the front glass or LCD. Be careful with
rings or other jewelry and do not apply excessive pressure on the f
ront glass or LCD.
CAUTION: Do not apply or spray li quid directly to the dis play as excess liquid may c ause
damage to internal electronics. Instead, apply the liquid to a cleaning cloth.
To clean t he display
Clean the display using a sponge, cleaning cloth or soft tissue, lightly moistened with a recognized clean-
ing product for medical equipment. Read and follow all label instructions on the cleaning product. In case
of doubt about a certain cleaning product, use plain water.
Possible cleaning solutions:
70% isopropyl alcohol
1.6% aqueous ammonia
Cidex® (2.4% glutaraldehyde solution)
Sodium hypochlorite (bleach) 10%
“Green soap (USP)
0.5% Chlorehexidine in 70% isopropyl alcohol
Like Cleansafe® optical cleaning liquid
Do not use following products:
K5902134 NIO COLOR 2MP 23" 05/07/2017