Operation
4.4.3 Messages at the start of a program
Blood/whole blood should not be heated if the current operating temperature of the device
is over 37 °C, for example if you have previously thawed and heated plasma and you then
activate a program for heating blood/whole blood. You can nevertheless still start the
"BLOOD" program.
Acknowledge the display with the
left-hand selection button ("Yes") to
start the program despite the water
temperature being too high.
Acknowledge the display with the
right-hand selection button ("No") if
you do not wish to start the program.
Figure 15: Message "Water temperature too high!"
4.4.4 Messages while a program is running
If the heating chamber cover is opened while a program is in progress, the remaining time
is 'frozen', a message appears in the display and heat transfer fluid circulation is paused.
An intermittent acoustic signal sounds if the cover is not closed within 16s.
Close the heating chamber cover to
resume the program.
Figure 16: Message "Close the cover!"
NOTE
If paddle motion was active when the cover was opened, it will not resume until the heat-
ing cushions have sufficiently filled again.
B L O OD
3 7 . 0 °C
W a t e r t e m pe r a t u re
to o h ig h !
= > 4 1 . 6° C
S t a r t t h e p r o g r am ?
Y e s N o
P L A S MA
C l o s e t h e co v e r !
R e m a i n in g t im e 1 0 :3 1
- S t o p +