Specifications
11 Specifications
Manufacturer Barkey GmbH & Co. KG
Gewerbestrasse 8
33818 Leopoldshoehe
Tel.: +49 (5202) 9801-0
Fax: +49 (5202) 9801-99
Type Barkey plasmatherm
Order No. 212.10043 230 VAC / 50 - 60 Hz
212.10044 115 VAC / 60 Hz
Typical capacity 4 bags / 500 ml or
2 bags / 800 ml
Emergency: 8 bags / 250 ml
Power supply 230 VAC / 50 - 60 Hz or 115 VAC / 60 Hz
Typical power consumption < 0.25 W Standby
< 35 Watt Program pause
< 75 W Program "PLASMA" 45°C
1600 Watt max.
Sound pressure level < 45 dB(A)
Current consumption 7 A at 230 VAC max.
14 A at 115 VAC max.
Fusing 8 A slow at 230 VAC
15 A slow at 115 VAC
Battery type Lithium CR 1225, 3 V
Interfaces/devices optional:
Barcode scanner
Log printer
Overtemperature protection software controlled: setpoint + 1.0 °C
electronic: 48 ± 1.0 °C
Temperature setting Variable from +37 to +45 °C
Default setting +37 °C
Display accuracy 0.1 °C
Operation Storage/transport
Ambient temperature +10 to +40 °C -20 to +70 °C
Humidity 30 to 75 % 30 to 90 %
(non condensing)
Air pressure 700 to 1060 hPa 700 to 1060 hPa
Operating mode Continuous