EasyManua.ls Logo

Barkey plasmatherm - Factory Program Configurations

Barkey plasmatherm
72 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Error messages
Error
number
Error description
Comments
10, 11
Display fault
There may not be any text visible in the display.
The display may not be correct.
20 to 23
Internal device error
30, 31
Checksum of the internal
device data is incorrect
32
Checksum of the program
memory is incorrect
33
Checksum of the data
memory is incorrect
40
Clock is incorrect
41
Error setting clock time
42
Error setting date
43
Error reading clock time
44
Error reading date
45
Clock not running
Replace battery
(by authorised specialists only).
Then set time and date.
60
The watchdog timer has
detected an error.
70 to 73
Error with temperature
measurement / temperature
sensors
Possible cause: sensor open circuit / short-
circuit
80 to 81
Overtemperature (soft-
ware) during temperature
measurement
Restart system / allow device to cool down
90
Pump and heater supply
voltage not available.
Possible cause: overtemperature
91
Faulty program function
92
Overtemperature (elec-
tronic) during temperature
measurement
Restart system / allow device to cool down
97
Faulty undulation function.
98
No power supply for pad-
dle motor.
Restart system / remove paddle blockages
100
Fill level in tank is critical;
heater has been switched
off
Restart system / top up the tank
110
Communication / interface
fault

Table of Contents

Related product manuals