System description
6
Operating Instructions Radiation sensors R3XX
Dimensions
R 310 / R311
R 312
R 320
Distance ratio
R 310 / R320
R 311
R 312
Transmitters
Input
For R 310, R 311, R 312, R 320
PT100 for ambient temperature (TR 41-10)
Interface
HART
®
protocol (FSK BELL 202, 1,2 kb/s)
Functions (configurable via HART
®
interface, with TR 41-10 also via keyboard)
Unit of measurement
°C or °F
Lower/upper range limits
0…2000 °C (32…3632 °F)
Emission factor
0.1…1
Transmission factor
0.1…1
Ambient temperature alarm
20...70 or 125 °C (68...158 or 257 °F), sensor-dependent
Damping
0...999.9 s
Maximum mode
0...999.9 s
Minimum mode
0...999.9 s
Fault current
3.9...21.5 mA
Fieldbus address
0...15 (0 = point to point, 1…15 = multidrop)
HART
®
address
0...15 (0 = point to point, 1…15 multidrop)
Display
LC-Display (TR 41-10)
Analogue output
Output signal
4...20 mA, linear
Permissible load
500 for standard version/U
H
= 24 V
Accuracy
Measuring accuracy
1 % of measuring range (at 23 °C and for ε = 1)
R 312: 1% above 50 °C object temperature, less than 50 °C 3%
Temperature sensitivity
0.03 %/C°