Figure 5.1.4. Timed recording interface.
5.1.4. Background recording
In background recording mode the SB records the peak values within a given time window. Please
note that event recording is not affected by the peak recording both services may run
simultaneously.
In the field Mode you choose what kind of peaks should be recorded:
Peak is the maximum value of the signal during each period (as defined below).
Peak + Vector Sum includes the maximum of the vector sum of the 3 channels
Peak + Dominant Frequency includes the dominant frequency for each period
Peak + Vector Sum + Dominant Frequency includes the vector sum and the dominant
frequency for each period
You have to set-up the Period and the total Duration of a file. For verification purposes, the SB
shows the number of periods in one file of the given duration as Number of samples in file.
Figure 5.1.5. Background recording interface.